Products

 

Title

Atom Syringe Pump S-1235

Product Photo
Product Photo
Four sensors check the syringe for proper attachment 
  to the unit. Accidents due to incorrect attachment are 
  thus prevented.

The pressure inside the infusion line is always 
  monitored and displayed as a trend graph. The 
  occlusion pressure level for alarm activation can 
  be selected in the adequate level.

A numeric key pad is used to input the flow rate 
  accurately and speedily.

A lithium-ion battery, which has no memory effect, is 
  used in the unit. The battery capacity will not 
  decrease even after repeated charging.

Operational errors are prevented by the messages 
  urging the user to check for correct operation at the 
  beginning of infusion or in priming.

Diprivan PFS 1%* is available.

Atom Syringe Pump S-1235

Power requirementsCustomer-specified
(Power consumption 15VA)
Built-in battery : Lithium-ion battery
Applicable syringeTerumo, Nipro, Top, JMS, B.Braun : 10,20,30,50mL (disposable)
B-D : 10,30,50mL (disposable)
Monoject : 50mL (disposable)
*Diprivan PFS 1% : 20,50mL (disposable)
AlarmsOcclusion, Low battery, Syringe off, Flange off, Plunger off, Clutch released, Syringe empty, Seif-diagnosis
AlertsNo Operation, Near empty
Special functionsAutomatic calculation of the flow rate of a medical fluid, Volume limit setting, History storage
Dimensions and weight32.5(W)×12.5(D)×12(H)cm  
Approx. 2.1kg
*Diprivan PFS 1% is a registered trademark of AstraZeneca Group(generic: propofol)

PAGE TOP